Research article
Validity of a Newly-Designed Rectilinear Stepping Ergometer Submaximal Exercise Test to Assess Cardiorespiratory Fitness
Rubin Zhang1,2, Likui Zhan3, Shaoming Sun3, , Wei Peng1,2, Yining Sun3  
Author Information
1 Institute of Intelligent Machines, Chinese Academy of Sciences, Hefei, Anhui, PR, China
2 Department of Automation, University of Science and Technology of China, Hefei, Anhui, PR China
3 Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei, Anhui, PR China

Shaoming Sun
✉ PhD Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei 230031, Anhui, PR China
Email: sunsm1978@outlook.com
Publish Date
Received: 06-03-2017
Accepted: 28-06-2017
Published (online): 08-08-2017
 
ABSTRACT

The maximum oxygen uptake (V̇O2 max), determined from graded maximal or submaximal exercise tests, is used to classify the cardiorespiratory fitness level of individuals. The purpose of this study was to examine the validity and reliability of the YMCA submaximal exercise test protocol performed on a newly-designed rectilinear stepping ergometer (RSE) that used up and down reciprocating vertical motion in place of conventional circular motion and giving precise measurement of workload, to determine V̇O2 max in young healthy male adults. Thirty-two young healthy male adults (32 males; age range: 20-35 years; height: 1.75 ± 0.05 m; weight: 67.5 ± 8.6 kg) firstly participated in a maximal-effort graded exercise test using a cycle ergometer (CE) to directly obtain measured V̇O2 max. Subjects then completed the progressive multistage test on the RSE beginning at 50W and including additional stages of 70, 90, 110, 130, and 150W, and the RSE YMCA submaximal test consisting of a workload increase every 3 minutes until the termination criterion was reached. A metabolic equation was derived from the RSE multistage exercise test to predict oxygen consumption (V̇O2) from power output (W) during the submaximal exercise test (V̇O2 (mL·min-1 )=12.4 ×W(watts)+3.5 mL·kg-1·min-1×M+160mL·min-1, R2= 0.91, standard error of the estimate (SEE) = 134.8mL·min-1). A high correlation was observed between the RSE YMCA estimated V̇O2 max and the CE measured V̇O2 max (r=0.87). The mean difference between estimated and measured V̇O2 max was 2.5 mL·kg-1·min-1, with an SEE of 3.55 mL·kg-1·min-1. The data suggest that the RSE YMCA submaximal exercise test is valid for predicting V̇O2 max in young healthy male adults. The findings show that the rectilinear stepping exercise is an effective submaximal exercise for predicting V̇O2 max. The newly-designed RSE may be potentially further developed as an alternative ergometer for assessing cardiorespiratory fitness and the promotion of personalized health interventions for health care professionals.

Key words: Maximum oxygen uptake, metabolic equation, rectilinear reciprocating motion, constant power output


           Key Points
  • The rectilinear stepping exercise is a simple modality of exercise, which requires only up and down movements of the legs. It overcomes the mechanical dead centers of circular motion and is mechanically efficient. It is potentially applicable to a large group of populations.
  • The RSE gives an accurate measurement of power output and ensures a constant power output independent of stepping cadence.
  • The RSE submaximal exercise test is valid and feasible for estimating V̇O2max in young healthy male adults compared with the CE maximal exercise test.
  • The rectilinear stepping exercise is an effective submaximal exercise mode for predicting V̇O2max. The RSE designed for this study may be potentially developed as a new and alternative ergometer to assess cardiorespiratory fitness and could be used in the future by healthcare professionals to promote personalized health interventions.

INTRODUCTION

Exercise physiologists suggest that direct measurement of maximal oxygen uptake (V̇O2 max) during graded maximal exercise is the criterion measure of cardiorespiratory fitness (Heyward, 2013). Nonetheless, this method requires expensive equipment and experienced personnel and many individuals are not able to meet the criteria for maximal exertion because of fatigue, low motivation, or physical limitations (Herda et al., 2014; Noonan et al., 2000). The submaximal exercise test estimates a person’s aerobic capacity based on the assumption that a linear relationship exists between oxygen uptake (V̇O2) and heart rate (HR) within a certain range of HR (Heyward, 2013). Compared with maximal exercise testing, a submaximal exercise test is safer and more appropriate in public health institutions and clinical practice when assessing cardiorespiratory fitness of individuals (Herda et al., 2014; American College of Sports Medicine, 2013).

The use of traditional modes of testing in the prediction of V̇O2 max, such as with a treadmill or cycle ergometer (CE), may prove to be limited in some populations, including elderly people or deconditioned individuals (Siconolfi et al., 1982). To enable exercise testing across a greater range of population groups, tests with some new modalities of ergometers have been designed. In a study presented by Billinger et al. (2012), the YMCA submaximal cycling protocol was adapted for use with the NuStep T5xr recumbent stepper and used to estimate VO2 peak in healthy adults (r = 0.91; SEE = 4.09 mL·kg-1·min-1). With the popularity and continued interest in step aerobic training, the researchers provided a relatively more accurate estimate of V̇O2 max for young women (20–25 years) who used this device for aerobic training (r = 0.57; SEE = 5.3mL·kg-1·min-1) as compared to estimates for their untrained counterparts (r = 0.00; SEE = 6.7 mL·kg-1·min-1) (Roy et al., 2004). Paré et al. (1993) validated the prediction of VO2 peak from a submaximal exercise test performed by paraplegics on a wheelchair ergometer (r = 0.83; SEE = 0.19L·min-1) and observed the linear relationship between V̇O2 and power output (W) (V̇O2= 0.79+0.02 W; r = 0.80, SEE = 0.22L·min-1). The design of these different modalities of exercise devices provides the publics and health professionals with alternative exercise modes and methods for assessing cardiorespiratory fitness.

In addition to these different exercise modalities, rectilinear reciprocating motion is seldom used as a regular exercising modality compared with circular reciprocating motion, such as cycle ergometer. When using a rotary crank mechanism, the force application, however, is not constant throughout the pedal-crank revolution (Ratel et al., 2004). Ericson and Nisell (1988) showed that the tangential force (perpendicular to the crank) exertion was maximal when the position of the crank was approximately horizontal (90 degree crank angle), and the tangential force was minimal at two points, known as the top dead center (0 degree crank angle) and bottom dead center (180 degree crank angle) (Chen et al., 1997; 2001; Ericson and Nisell, 1988; Patterson and Moreno, 1990). Moreover, Ericson and Nisell (1988) found that the tangential force was the only mechanically efficient force while the centrifugal force (perpendicular to the tangential force and running alongside the crank) did not contribute to the end mechanical efficiency. Optimum mechanical efficiency was reached if the centrifugal force equaled zero, implying that the sum of forces was directed tangentially and acted in the same instantaneous direction as the pedal moves (Ericson and Nisell, 1988). Therefore, a mechanism of only linear vertical motion of the pedals will overcome the dead centers and be mechanically efficient because the vertical force is nearly the only tangential force and it is entirely in the direction of the pedal’s movement. Since some individuals are unfamiliar with cycling exercise, a rectilinear stepping motion using only linear vertical motion of the lower extremities, may provide a simple and alternative mode of exercise for them. In addition, Miller et al. (2001) suggested that a stepper cycle powered by a reciprocating vertical motion was applicable for people with cognitive, orthopedic, and neuromuscular conditions as it reduced the required range of movement of the lower limb joints when compared to a conventional bicycle. Moreover, many different types of stepping activities, including using various types of step ergometers, were found to be beneficial to knee muscle strengthening and rehabilitation programmes (Willett et al., 2010).

Our research team designed a new rectilinear stepping ergometer (RSE) that was based on a spinning bike and used up and down reciprocating vertical motion of the lower extremities in place of conventional circular motion. The purpose of this study was to examine the validity and reliability of the YMCA submaximal exercise test protocol for determining V̇O2 max in young healthy male adults using the RSE. We hypothesized that the RSE YMCA submaximal exercise test would be a valid test for determining V̇O2 max.

METHODS

Subjects

Thirty-two healthy males (n = 32; Age range: 20-35 years; Height: 1.75 ± 0.05 m; Weight: 67.5 ± 8.6 kg) were recruited to participate in this study. Subjects’ age and physical characteristics are presented in Table 1. All subjects were screened and free from major cardiovascular risk, orthopedic and metabolic diseases. All procedures in this study were approved by our local ethics committee. Before testing, all subjects were asked to sign a written institutionally approved informed consent and they were informed of the benefits and risks of the study prior to signing the document. The subjects performed a CE maximal exercise test and an RSE submaximal exercise test to test the validity of the RSE for predicting V̇O2max. Reliability testing consisted of two paired test-retest sessions separated by no more than 4 days. All tests were separated by at least 48 hours and completed within 3 weeks. Before each test, subjects were asked to avoid alcohol and any heavy exercise for at least 6 hours. Subjects were familiar with the CE, but not all were familiar with RSE and every subject had an opportunity to practice the rectilinear stepping movement pattern and step cadence for approximately 15 minutes with no load prior to the RSE exercise test.

Instrumentation

The rectilinear stepping ergometer (RSE) (Figure 1) used for the present study was designed by members of the Research Center for Information Technology of Sports and Health. A high-accuracy pressure and displacement sensor were respectively mounted on the pedals and the rectilinear motion guiding unit, to accurately measure the force and speed in the vertical direction. It is simple for giving direct measurement of power output on the RSE by multiplying the measured force by speed (power = force × speed). An electric generator braking system was connected to the rectilinear motion guiding unit through a gear transmission unit. The vertical motion of pedals was transformed into the circular motion of the gears through a rectilinear motion guiding unit and actuated the rotation of the generator rotor to produce the brake power through a gear transmission unit. Braking power ranged from 30 up to 400 watts. The resistance of RSE could be instantly adjusted through the change of exciting current of the electric generator, which kept a constant power output (accuracy ± 1.5 watts) with a stepping cadence over a range of 80-120 steps per minute. When instant power output exceeded the target value, the exciting current decreased to lower the resistance and the instant power output decreased simultaneously. Conversely, the exciting current was increased when the instant power output was under the target value. This has been found to be important during submaximal exercise testing because stepping cadence fluctuates and maintaining constant power allows subjects to reach a steady heart rate during each stage of the test (Herda et al., 2014; Billinger et al., 2012). The pedal stroke length of circular crank pedals cannot be adjusted, but the pedal stroke length of the RSE could be changed through an adjustable limit switch to fit the user’s leg and stroke pattern, allowing short or long strokes. The height of the saddle and the handles could be adjusted manually for maximum comfort.

Procedures

Maximum cycle ergometer exercise test

Subjects participated in a cycle ergometer graded maximal exercise test, which consisted of a 4-minute warm-up phase of pedaling with no load, followed by a progressive phase with workload increments of 25 W every minute (Maud et al., 1995). Subjects were required to maintain the cadence at fixed 60 rpm throughout the test. The test was terminated when at least three of the following criteria were achieved: (a) the subject reached volitional fatigue and requested to end the test (b) achievement of an age-predicted maximal HR (220-age) (c) a plateau in oxygen consumption with increasing workload (d) the subject was unable to maintain the cadence (e) a respiratory exchange ratio (RER) greater than 1.10.

RSE multistage exercise test

Before using the submaximal exercise test to predict V̇O2 max, a metabolic equation describing the relationship between oxygen consumption (V̇O2) and power output (W) adapted for RSE was needed to be developed to predict V̇O2 from W. Subjects were asked to perform a progressive multistage exercise test and maintain the cadence at 100 steps·min-1 throughout the test (Figure 2). After a 3 minutes warm-up with an initial workload of 30W, the test began at 50 W and included additional stages of 70, 90, 110, 130, and 150 W. Steady rate V̇O2 at each stage was recorded, which occurred when the absolute difference in oxygen consumption between two successive minutes after 3 min of the current stage did not exceed 50 mL·min-1, and the mean of those two minutes was used as the steady rate value. After reaching a steady state, the workload was increased to the next stage until the subjects reached 85% of the age-predicted maximal HR (220 - age) or finished the 150 W stage. On completion of the progressive multistage exercise test, a metabolic equation to predict V̇O2 from W using the RSE was developed using linear regression analysis.

RSE submaximal exercise test

After completing the CE maximal exercise test, subjects were asked to perform a YMCA submaximal exercise test on the RSE. They were required to maintain the cadence at 100steps·min-1 throughout the test. The YMCA protocol (Table 2) adapted for the RSE started with an initial workload at 50 W and the resistance was increased every 3 minutes according to the protocol (Table 2) until one of the following criteria was achieved: (a) volitional fatigue (b) 85% of the age-predicted maximal HR (220 - age) (c) submaximal test was completed. Each stage lasted 3 minutes and an extra minute was added to obtain the steady-state HR if HR varied by more than 5 beats·min-1 between the second and third minutes of each stage. If the variation was within 5 beats·min-1, the steady-state HR of the current stage was recorded and subjects were asked to progress to the next stage. The test assumed a linear relationship between HR and workload at 110-150 beats·min-1 HR range (Golding et al., 1989). HR collected from the last two stages of the test and the respective workload calculation were used to draw a straight line between workload (x - axis) and HR (y - axis) and the line extended to the subject’s age-predicted maximal HR (220 - age) to calculate the corresponding maximal working capacity (Wpeak). V̇O2 max was subsequently calculated using the RSE metabolic equation which was developed previously.

Data collection

Data collection was performed at the Research Center for Information Technology of Sports and Health. The cycle ergometer and RSE exercise test were conducted separately on a Monark cycle ergometer (Monark 839E, Sweden) and the RSE invented by members of the Research Center for Information Technology of Sports and Health. The seat and handle heights/positions of both sets of equipment were adjusted according to each subject’s comfort. A metabolic measurement system (MedGraphics VO2000, UK) was used to measure and analyze oxygen uptake through collection of expired gases during the tests. Subjects wore a nose clip and expired through a two-way rebreathing valve connected to the metabolic cart. HR was continuously monitored throughout each exercise testing session using a Polar HR monitor (Polar Electro, Finland). All equipment was calibrated precisely before testing according to the manufacturer’s recommendations.

Statistical analysis

The arithmetic mean and standard deviation were used for descriptive statistics. Pearson correlation coefficient (r) was used to determine the relationship between the CE V̇O2 max test and the RSE YMCA submaximal test and a Bland-Altman plot was constructed to determine the level of agreement between the CE measured V̇O2 max and RSE YMCA estimated V̇O2 max. The standard error of the estimate was calculated to estimate the accuracy of the RSE YMCA estimated V̇O2 max and CE measured V̇O2 max. The mean difference of the prediction was calculated by subtracting the RSE YMCA V̇O2 max test result from the CE V̇O2 max test result. Paired t test was used to examine for a significant difference between the two test results. To determine the test-retest reliability for each testing protocol, the intra-class correlation coefficients (ICC) were calculated for the CE maximal testing and RSE submaximal testing protocols. Regression analysis was used to develop the equation to predict oxygen uptake (V̇O2) from power output (W).

RESULTS

All subjects’ individual data points and a regression line of measured oxygen uptake (V̇O2) on power output (W) from the RSE multistage test are presented (Figure 3). The linear equation (A) was represented to calculate V̇O2 from W for the RSE test and a high correlation was observed between the two variables (R2 = 0.88):


Similarly to the ACSM leg ergometry metabolic equation (Heyward, 2013), after factoring in an estimation of resting metabolism per kg of body weight (M) the additional unloaded stepping V̇O2was approximately 160 mL·min-1, the final equation (B) was developed and then used for predicting V̇O2 from a submaximal power output (W).


Mean and standard deviation values for the two exercise test (CE maximal and RSE submaximal exercise test) variables are reported (Table 3). The mean difference between two variables was 2.5mL·kg-1·min-1. A paired t test revealed no significant difference between CE V̇O2 max and RSE YMCA V̇O2 max (p < 0.01). Pearson’s correlation coefficient between the RSE YMCA submaximal exercise test for estimated V̇O2 max and CE maximal exercise test for measured V̇O2 max (r = 0.87, Figure 4) indicated a strong relationship between the two tests for assessing cardiorespiratory fitness in young healthy male adults, and the standard error of the estimate (SEE) was 3.55 mL·kg-1·min-1. The Bland-Altman plot (Figure 5) presents the agreement between CE V̇O2 max and RSE YMCA V̇O2 max. The bias for two V̇O2 max values was 2.5mL·kg-1·min-1, with limits of agreement of -4.3mL·kg-1·min-1 and 9.3mL·kg-1·min-1. High test-retest reliabilities were found for both the CE maximal exercise test (ICC = 0.91) and the RSE submaximal exercise test (ICC = 0.92).

DISCUSSION

Since laboratory direct measurement of V̇O2 max is expensive, time-consuming and requires well-trained personnel, submaximal exercise testing is a more practical and efficient way in clinical practice when assessing cardiorespiratory fitness. The prediction of V̇O2 max using submaximal testing is more appropriate in certain venues such as fitness centers, rehabilitation clinics, and assisted living facilities (Billinger et al., 2012), where maximal testing is not feasible. In this study, we designed a rectilinear stepping ergometer (RSE) as a new modality of exercise device for assessing cardiorespiratory fitness and our findings suggested that the YMCA submaximal exercise test protocol performed on the RSE could accurately predict V̇O2 max in young healthy male adults. The study potentially provides the publics and health care professionals a new and alternative modality of exercise to assess cardiorespiratory fitness.

Our findings suggested that the RSE YMCA submaximal exercise test was valid and feasible for estimating V̇O2 max in young healthy male adults. The RSE YMCA submaximal exercise test was strongly correlated with the CE maximal exercise test for assessing V̇O2 max (r = 0.87) and the mean difference between the two tests was 2.5 mL·kg-1·min-1, which was not significantly different (p < 0.01). The SEE reflects the accuracy of the estimate, a lower SEE value indicating a greater accuracy of the predicted result. In our study, the SEE of 3.55 mL·kg-1·min-1 was close to 1 metabolic equivalent (1 MET = 3.5 mL·kg-1·min-1) and considered appropriate for predicting V̇O2 max (Heyward, 2013). Furthermore, as illustrated in Figure 5, the limits of agreement presented that both upper and lower limit were below the recommended 10 mL·kg-1·min-1 (Heyward, 2013), which reflects a good predictive capability.

Although the mean difference between CE V̇O2 max and RSE YMCA V̇O2 max was no more than 3 mL·kg-1·min-1, we found that the RSE YMCA submaximal exercise test did overpredict V̇O2 max, which was 2.5 mL·kg-1·min-1 higher than CE measured V̇O2 max. A previous cross-validated study by Beekley et al. (2004) reported the validity of the CE YMCA test against the maximal exercise test of CE and treadmill (TM). They found a moderately high significant correlation between TM V̇O2 max and predicted V̇O2 max from the CE YMCA test (r = 0.77, p < 0.05), with a mean difference between the two variables of 1.84 mL·kg-1·min-1, indicating that the current CE YMCA submaximal test appeared to be effective in predicting TM V̇O2 max rather than CE V̇O2 max. Beekley et al. (2004) suggested some factors that may have led to a prediction error between the TM measured V̇O2 max and the CE YMCA predicted V̇O2 max, including the sex of the subjects, differences in exercise modalities, and only partial linearity between the power output (W) and heart rate (HR) in the YMCA protocol.

In our study, the prediction difference between the two exercise tests may be attributed to several factors. First, the RSE metabolic equation was developed based on the relationship between actual V̇O2 and W and we used this equation to predict V̇O2 from W during the submaximal test. The slope of our developed equation was 12.4 mL·watt-1·min-1, which was slightly higher than the slope of the current ACSM leg ergometer metabolic equation (10.8 mL·watt-1·min-1) (Heyward, 2013). This increase of slope could be explained by the use of different exercise modalities and may result in a higher value for the RSE estimated V̇O2 max. Lang et al. (1992) mentioned that the intercept of the metabolic equation, including a resting metabolism estimation and an extra V̇O2, could be another important source of prediction error. Second, the YMCA submaximal test was based on the assumption that the linear relationship existed between W and HR (Golding et al., 1989). However, several studies have mentioned that this linearity existed only at certain heart rates (Beekley et al., 2004; Golding et al., 1989; Lang et al., 1992). As described in the Y’s way to Physical Fitness (Golding et al., 1989), at low heart rates, many external stimuli, such as talking, laughter, and nervousness, could affect the heart rate. Also, when exercise intensity exceeded the gas exchange threshold (GET), the relationship was no longer linear (Jamnick et al., 2016; Londeree et al., 1997). Thus, in our study, this nonlinearity may result in an overprediction of V̇O2 max. In addition, the exercise modality could have a crucial influence on prediction difference. Despite both the CE and RSE belonging to lower-extremity exercise, the muscle groups mobilized in these two exercises are different and this may be an important factor in prediction accuracy.

To our knowledge, our study is the first to study exercise testing using a rectilinear reciprocating motion. We chose to study this rectilinear stepping exercise modality for several reasons. First, compared with traditional circular motion, the up and down rectilinear motion overcomes the mechanical dead centers, known as the “dead spots” in some studies (Chen et al., 1997; Chen et al., 2001). Chen et al. (1997) described the “dead spots” at 0 and 180 degree of crank arm position in cycling with functional electrical stimulation (FES), basically the transition point between knee extension and flexion during the cyclical movement of the leg. Other researchers have reported that, under a stable cycling status, the cycling speed decreased more violently at the mechanical “dead spots” angle at which points the stimulated muscle has minimal amount of torque output (Chen et al., 2001). Moreover, Ericson and Nisell (1988) suggested that in rotary crank mechanism, the optimum mechanical efficiency was reached if the centrifugal force equaled zero, implying that the sum of forces was directed tangentially and acted in its entirety in the direction as the pedal moves. Therefore, we suggest that such a rectilinear reciprocating motion of RSE is mechanically efficient because the vertical force is nearly the only tangential force and it is entirely in the direction of the pedal’s movement. Second, since some individuals are unfamiliar with cycling exercise (American College of Sports Medicine, 2013), the RSE maybe more simple for them to use. A short-stroke stepping pattern of the RSE may be applicable for those who cannot perform large scale leg movement, such as elderly people or individuals with knee injury. Additionally, RSE is a lower-limb exercise modality, and lower-limb dominant exercise training improves flow-mediated vasodilation (FMD) in the lower extremities for patients with chronic heart failure (CHF) (Kobayashi, et al., 2003).

Another advantage of RSE is that a constant power output is ensured with stepping cadence fluctuating over certain ranges. The exciting current of the electric generator instantaneously changes to adjust the resistance based on the proportion integration differentiation (PID) control principle, to inversely change the increase or decrease of power output (W) caused by the fluctuation of stepping cadence. When instant power output exceeds the target value, the exciting current needs to decrease to lower the resistance, so the instant power output decreases simultaneously. Conversely, the exciting current will be increased when the instant power output is under the target value. Thus, the constant power output is maintained during the exercise test. This is crucial for standardizing the workload across all subjects and maintaining constant power allows subjects to reach a steady state HR (within 5 beats·min-1) during submaximal exercise test (Billinger et al., 2012; Herda et al., 2014).

The results presented in this study are limited to a relatively small sample size (n = 32) of male subjects. The validity of the RSE YMCA submaximal exercise test for estimating V̇O2 max needs to be confirmed in a wider sample including females and subjects of different ages and physical condition. In addition, the RSE metabolic equation should be revised by considering gender and stepping cadence as determinants of V̇O2 and its validity requires testing on a larger variety of populations. Another limitation of the present study is that the volunteers we recruited had a predetermined positive attitude toward the study and physical activity. Subjects unaccustomed to the exercise on the RSE may reach volitional fatigue more quickly due to leg muscle fatigue similar to that of a cycle ergometer, affecting the prediction accuracy of V̇O2 max.

There is a need for further research to improve accuracy in the prediction of V̇O2 max from the RSE submaximal exercise test with a larger variety of populations, and to examine the validity of the RSE graded maximal exercise test in determiningV̇O2 max. Furthermore, research into the biomechanical advantages and efficiency of energy expenditure from this novel exercise modality would also be of value.

CONCLUSION

A rectilinear stepping exercise modality was presented in this study and the results indicated that the RSE submaximal exercise test was valid and reliable for estimating V̇O2 max in young healthy male adults. Since submaximal exercise test is more practical in assessing cardiorespiratory fitness without expense, risk, and hard effort, the result of this study will provide the rectilinear stepping exercise as an effective submaximal exercise mode for predicting V̇O2 max for the publics. The RSE designed for this study may be potentially developed as a new and alternative ergometer to assess cardiorespiratory fitness and could be used in the future by healthcare professionals to promote personalized health interventions.

ACKNOWLEDGEMENTS

The study was supported by the Science and Technology Development Program of Anhui Province of China (NO.1604a0902188), the Science and Technology Major Program of Anhui Province of China (NO.15czz02072), and the Science and Technology Service Network Program, Chinese Academy of Sciences (NO. KFJ-SW-STS-161). We thank the staff and students of the Research Center for Information Technology of Sports and Health for their assistance in data collection, and the volunteer participants. There are no financial or other relationships that could have led to a conflict of interest. The experiments performed in this study comply with the current laws of China.

AUTHOR BIOGRAPHY

Journal of Sports Science and Medicine Rubin Zhang
Employment: Institute of Intelligent Machines, Chinese Academy of Sciences
Degree: PhD
Research interests: Exercise physiology
E-mail: zrb1990@mail.ustc.edu.cn
 

Journal of Sports Science and Medicine Likui Zhan
Employment: Hefei Institutes of Physical Science, Chinese Academy of Sciences
Degree: MS
Research interests: Exercise physiology, Information acquisition
E-mail: zlk1968@outlook.com
 

Journal of Sports Science and Medicine Shaoming Sun
Employment: Hefei Institutes of Physical Science, Chinese Academy of Sciences
Degree: PhD
Research interests: Exercise physiology
E-mail: sunsm1978@outlook.com
 

Journal of Sports Science and Medicine Wei Peng
Employment: Institute of Intelligent Machines, Chinese Academy of Sciences
Degree: MS, PhD student
Research interests: Exercise rehabilitation, Information acquisition
E-mail: wpeng@mail.ustc.edu.cn
 

Journal of Sports Science and Medicine Yining Sun
Employment: Hefei Institutes of Physical Science, Chinese Academy of Sciences
Degree: PhD
Research interests: Exercise physiology
E-mail: syn1963@outlook.com
 
 
REFERENCES
Journal of Sports Science and Medicine Beekley M.D., Brechue W.F., Dehoyos D.V., Garzarella L., Werber-Zion G., Pollock M.L. (2004) Cross-validation of the YMCA submaximal cycle ergometer test to predict V. Research Quarterly for Exercise & Sport 75, 337-342.
Journal of Sports Science and Medicine Billinger S.A., Van S.E., McClain M., Lentz A.A., Good M.B. (2012) Recumbent stepper submaximal exercise test to predict peak oxygen uptake. Medicine & Science in Sports & Exercise 44, 1539-1544.
Journal of Sports Science and Medicine Chen H.Y., Yu N.Y., Chen K., Tsai K.H., Fu L., Chen S.C. (2001) Development of FES-cycling System with Network Capability for Multi-Center Clinical Studies. Journal of Medical & Biological Engineering 21, 85-92.
Journal of Sports Science and Medicine Chen J.J., Yu N.Y., Huang D.G., Ann B.T., Chang G.C. (1997) Applying fuzzy logic to control cycling movement induced by functional electrical stimulation. IEEE Transactions on Rehabilitation Engineering 5, 158-169.
Journal of Sports Science and Medicine Damiano D.L., Norman T., Stanley C.J., Park H.S. (2011) Comparison of elliptical training, stationary cycling, treadmill walking and overground walking. Gait & Posture 34, 260-264.
Journal of Sports Science and Medicine Ericson M.O., Nisell R. (1988) Efficiency of pedal forces during ergometer cycling. International Journal of Sports Medicine 9, 118-122.
Journal of Sports Science and Medicine Formenti F., Minetti A.E., Borrani F. (2015) Pedaling rate is an important determinant of human oxygen uptake during exercise on the cycle ergometer. Physiological Reports 3, e12500-.
Journal of Sports Science and Medicine Glaner M.F., Silva R.A. (2011) Validation of a new cycle ergometer. International Journal of Sports Medicine 32, 117-121.
Journal of Sports Science and Medicine Golding L.A., Myers C.R., Sinning W.E. (1989) Y’s way to physical fitness: the complete guide to fitness testing and instruction. Human Kinetics Publishers.
Journal of Sports Science and Medicine Herda A.A., Lentz A.A., Mattlage A.E., Sisante J.F., Billinger S.A. (2014) Cross-validation of the recumbent stepper submaximal exercise test to predict peak oxygen uptake in older adults. Phys Ther 94, 722-729.
Journal of Sports Science and Medicine Heyward V (2013) Advanced Fitness Assessment and Exercise Prescription. Champaign, IL. Human Kinetics Inc..
Journal of Sports Science and Medicine Jamnick N.A., By S., Pettitt C.D., Pettitt R.W. (2016) Comparison of the YMCA and a Custom Submaximal Exercise Test for Determining V. Medicine & Science in Sports & Exercise 48, 254-259.
Journal of Sports Science and Medicine Kobayashi N., Tsuruya Y., Iwasawa T., Ikeda N., Hashimoto S., Yasu T., Ueba H., Kubo N., Fujii M., Kawakami M., Saito M. (2003) Exercise training in patients with chronic heart failure improves endothelial function predominantly in the trained extremities. Circulation Journal Official Journal of the Japanese Circulation Society 67, 505-510.
Journal of Sports Science and Medicine Lang P.B., Latin R.W., Berg K.E., Mellion M.B. (1992) The accuracy of the ACSM cycle ergometry equation. Medicine & Science in Sports & Exercise 24, 272-726.
Journal of Sports Science and Medicine Londeree B., Moffitt-Gerstenberger J., Padfield J.A., Lottmann D. (1997) Oxygen consumption of cycle ergometry is nonlinearly related to work rate and pedal rate. Medicine & Science in Sports & Exercise 29, 775-780.
Journal of Sports Science and Medicine Maud P.J., Foster C. (1995) Physiological assessment of human fitness. Human Kinetics.
Journal of Sports Science and Medicine Miller M.S., Peach J.P., Keller T.S. (2001) Electromyographic analysis of a human powered stepper cycle during seated and standing riding. Journal of Electromyography & Kinesiology Official Journal of the International Society of Electrophysiological Kinesiology 11, 413-423.
Journal of Sports Science and Medicine Nakamura K., Ohira M., Yokokawa Y., Nagasawa Y. (2015) Validity and reproducibility of an incremental sit-to-stand exercise test for evaluating anaerobic threshold in young, healthy individuals. Journal of Sports Science & Medicine 14, 708-715.
Journal of Sports Science and Medicine Noonan V., Dean E. (2000) Submaximal exercise testing: clinical application and interpretation. Physical Therapy 80, 782-807.
Journal of Sports Science and Medicine Paré G., Noreau L., Simard C. (1993) Prediction of maximal aerobic power from a submaximal exercise test performed by paraplegics on a wheelchair ergometer. Paraplegia 31, 584-592.
Journal of Sports Science and Medicine Patterson R.P., Moreno M.I. (1990) Bicycle pedaling forces as a function of pedaling rate and power output. Medicine & Science in Sports & Exercise 22, 512-516.
Journal of Sports Science and Medicine Ratel S., Duché P., Hautier C.A., Williams C.A., Bedu M. (2004) Physiological responses during cycling with noncircular “harmonic” and circular chainrings. European Journal of Applied Physiology 91, 100-104.
Journal of Sports Science and Medicine Roy L.P., Smith J.F., Bishop P.A., Hallinan C., Wang M., Hunter G.R. (2004) Prediction of maximal V from a submaximal StairMaster test in young Women. Journal of Strength and Conditioning Research 18, 92-96.
Journal of Sports Science and Medicine Siconolfi S.F., Cullinane E.M., Carleton R.A., Thompson P.D. (1982) Assessing V max in epidemiologic studies: modification of the Astrand-Rhyming test. Medicine & Science in Sports & Exercise 14, 335-338.
Journal of Sports Science and Medicine Storer T.W., Davis J.A., Caiozzo V.J. (1990) Accurate prediction of V in cycle ergometry. Medicine & Science in Sports & Exercise 22, 704-712.
Journal of Sports Science and Medicine Willett G.M., Karst G.M., Canney E.M., Gallant D., Wees J.M. (2010) Lower limb emg activity during selected stepping exercises. Journal of Sport Rehabilitation 7, 102-111.
Journal of Sports Science and MedicineAmerican College of Sports Medicine. (2013) Interpretation of Results. In: ACSM’s Guidelines for Exercise Testing and Prescription. Eds: Pescatello, L.S., Arena, R., Riebe, D., Thompson, P.D. 9th edition. Philadelphia: Lippincott Williams & Wilkins. 87.
 
 
 
Home Issues About Authors
Contact Current Editorial board Authors instructions
Email alerts In Press Mission For Reviewers
Archive Scope
Supplements Statistics
Most Read Articles
  Most Cited Articles
 
  
 
JSSM | Copyright 2001-2024 | All rights reserved. | LEGAL NOTICES | Publisher

It is forbidden the total or partial reproduction of this web site and the published materials, the treatment of its database, any kind of transition and for any means, either electronic, mechanic or other methods, without the previous written permission of the JSSM.

This work is licensed under a Creative Commons License Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.